| Home Page
About Us
Induction Loop Systems
Induction Loop Receivers
Induction Loop Testing Equipment
Ambient Noise Compensation
Sound Field System Accessory
Multisensory Sound Lab
Airplane ALS
Client List
ALS Retail Pricing
Contact Us |
Standards:
Induction Loop Assistive Listening Systems
Please Note: An accurate field strength meter such as the FSM
Induction Loop Test Set is required to check compliance with IEC 118-4
Technical Standards.
Summary of IEC 118-4 (1981) Standards:
With a signal of 1,000 Hertz at a level equal to the long-time average
of the speech signal applied to the loop system's input, the resulting
field strength inside the area(s) enclosed by the loop(s) shall average
100 mA/meter +/- 3 dB, with 70 mA/meter as the minimum level and
140 mA/meter as the maximum level, measured 1.2 meters above the floor.
Allowing for 12 dB peaks occurring in speech, peak field strength may reach
400 mA/meter. The frequency response of the induction loop system
shall be 100 Hz-5 kHz +/- 3 dB.
Additional Specifications Suggested by Oval Window
Audio (rev 8/03):
1. Loop Wire Installation: In order to maximize signal strength
and uniformity, the loop wire shall be installed either at floor level
or at a ceiling height not to exceed the manufacturer's recommendations.
At least 80% of the installed loop wire should be free of the influences
of metal, either in front of, or immediately behind the wire.
2. Field Strength:
Pure tone: As per IEC 118-4 with the additional conditions that
the measurement be "A" weighted to disallow the influence of inaudible
low frequency power line electrical noise, and that the measurement at
the center of the loop(s) be 100 mA/meter +/- 3 dB at a height of 1.2 meters,
with an input signal of 1,000 Hertz.
Speech: Using unweighted/flat frequency response setting of
measurement equipment, speech test signals must average + 3 dB at
the center and throughout the listening area, with signal peaks not exceeding
+12 dB re: 100ma/meter.
3. Input Signal Compression: To achieve a long-time average
field strength of 100 mA/meter, and to maximize
signal intelligibility and listener comfort regardless of fluctuating signal
input levels, an automatic gain control, signal compression and/or adjustable
non-distorting peak limiter must be employed at the input of the system.
Recommended compression ratios: 4:1 (music), up to 20:1 (speech).
4. Frequency Response: The frequency response of an induction
loop system shall comply with IEC 118-4 specifications, as confirmed with
a field strength meter. Frequency response measurements conducted
with an "A" weighted field strength meter will need to be corrected to
compensate for the substantial low frequency roll off characteristic of
this weighting. The use of a graphic equalizer is beneficial in certain
situations where the environment is affecting the system's frequency response,
and/or the listening audience consists of people who have specific needs
and/or preferences.
5. Ambient Electrical Interference: Sources of electromagnetic
radiation that may interfere with the proper functioning of an induction
loop system include: faulty florescent lighting, light dimmers, electrical
wiring, TV and computer monitors, surge protectors and radio transmitters
that are in close proximity to the loop system. When possible, an
on-site evaluation of ambient electromagnetic noise should be performed
before a loop system is installed in order to identify and resolve any
potential sources of interference. Using a field strength meter,
ambient electromagnetic noise should not exceed -20 dB ("A" weighted) re:
100 mA/meter, as measured at any seat within the area that is to be enclosed
by the loop. Depending on the location, intensity and frequency
of the electrical noise, and the potential for remedying the problem, the
use of an induction loop system may have to be ruled out. Using an
induction receiver or telecoil equipped hearing aid, listening tests may
be required to determine if the frequency(ies) of the electrical noise
fall within the frequency response range of hearing aids. For example,
induction receivers and most hearing aids do not reproduce sound below
100 Hertz. As a result, 50/60 Hz. power line noise should not cause
a problem, even though it may be measurable on some field strength meters,
unless higher (audible) harmonics of the noise are present.
6. Signal Spillover: In the event adjacent areas are equipped
with induction loop systems, signal spillover must not exceed 12.5 mA/meter,
or -18 dB ("A" weighted), re: 100 mA/meter, as measured at
any location within any of the induction loop(s) in use. Using an
induction receiver or telecoil equipped hearing aid, listening tests should
be performed with all systems in use in order to determine if the signal
spillover is audible and/or distracting.
7. System Signal-to-Noise: The electrical signal to noise
ratio of the loop amplifier output (measured directly, not inductively)
at 1,000 Hz. must be at least +30 dB (unweighted) re: to the system's internal
noise at an output level sufficient to deliver 100 mA/meter as per IEC
118-4 specifications.
8. Distortion: With an input signal of 1,000 Hz and the
system adjusted for an output of 100 mA/meter, harmonic distortion must
not exceed 3%.
9. Listening Tests: a regularly scheduled listening check
of all assistive listening systems and their
associated receivers is strongly recommended, preferably before each use.
FOR ADDITIONAL INFORMATION ON ANY OF OUR INDUCTION
LOOP SYSTEMS, PLEASE CONTACT US BY E-MAIL,
OR PHONE/FAX 303-447-3607.
Home Page // About
Us // Induction Loop Assistive Listening
Systems // Induction Loop Receivers
// Induction Loop Testing Equipment //
Ambient
Noise Compensation //
Sound Field System
Accessory // Vibrotactile/Multisensory
Sound Lab //
Airplane Assistive Listening
System // Client List //
ALS Pricing & Ordering // Contact Us
This site created with the help of The
Write Direction, Kauai, HI.
This
site is optimized for viewing with Netscape 3 or higher. To get Netscape,
click on Netscape Now!
|